Calibration – HP E1418A User Manual
Page 76
![background image](/manuals/90248/76/background.png)
•
This query returns one of the following values:
Return
Value
Meaning
0
Calibration is complete.
New constants are in RAM.
1
More measurements are needed.
5
Calibration aborted.
•
A return value of 5 (calibration aborted) is given if either:
a. The command is received when the module is in current calibration or
resistance calibration mode, or
b. The calibration in process has entered more than 50 values without a
successful calibration.
•
When the channel voltage calibration is completed, the new calibration constants
are used by the module, but not stored in non-volatile memory.
•
If a voltage calibration is interrupted before completion (that is, by another
command being received), no constants are changed.
•
Refer to Appendix D, “ Voltage/Current Output Adjustment” , starting on page 154,
for additional information about this command.
•
Related Commands:
CALibration:STORe
,
CALibration
n
:CONFigure:VOLTage?
CALibration
Chapter 3
HP E1418A SCPI Command Reference 75