beautypg.com

Calibration isolation status register – HP E1418A User Manual

Page 147

background image

Calibration

Isolation Status

Register

The Calibration Isolation Status Register provides a non-volatile record of
the isolation status of each channel at the last calibration. This register is
non-volatile.

Address

Base + 188

16

Bit

15

14

13

12

11

10

09

08

07

06

05

04

03

02

01

00

READ

Ch

16

Iso

Cal

Stat

Ch

15

Iso

Cal

Stat

Ch

14

Iso

Cal

Stat

Ch

13

Iso

Cal

Stat

Ch

12

Iso

Cal

Stat

Ch

11

Iso

Cal

Stat

Ch

10

Iso

Cal

Stat

Ch

09

Iso

Cal

Stat

Ch

08

Iso

Cal

Stat

Ch

07

Iso

Cal

Stat

Ch

06

Iso

Cal

Stat

Ch

05

Iso

Cal

Stat

Ch

04

Iso

Cal

Stat

Ch

03

Iso

Cal

Stat

Ch

02

Iso

Cal

Stat

Ch

01

Iso

Cal

Stat

WRITE

Ch XX Iso Cal Stat: This field is set to a one (1) when the last calibration
of the channel was performed with the channel configured as non-isolated.
This field is set to a zero (0) when the channel was last calibrated in the
isolated configuration.

Notes

This register is not set or used by the hardware. The HP E1418A SCPI
driver, if used to calibrate the module, automatically sets the bits in this
register. If you use register programming to calibrate this module, you may
wish to use this register to record the calibration status.

The Voltage Calibration Status Register, the Current Calibration Status
Register, the Calibration Isolation Status Register, and the Calibration Card
Configuration Register are used together to maintain the overall calibration
status of the module.

Changes in the values of these fields indicate a change in the module
configuration and may indicate the need to recalibrate the module.

146 HP E1418A Register-Based Programming

Appendix B