beautypg.com

Source n :voltage[:level]:triggered[:amplitude, Source – HP E1418A User Manual

Page 94

background image

SOURce

n

:VOLTage[:LEVel]:TRIGgered[:AMPLitude]

SOURce

n

:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <

value

| MIN | MAX | DEF>

outputs the voltage set in value at channel n when the trigger conditions are satisfied
and the channel output relay is closed. Use this command to synchronize multiple
channel outputs.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

SOURce

n

keyword

1 to 8 or 1 to 16

1

<

value

>

numeric (float 32) or

discrete

-16 V to +16 V or

MIN | MAX | DEF

0

Comments

In the keyword,

SOURce

n, the valid range of n is:

1 to 8 for

E1418A, E1418A Option 002

1 to 16 for

E1418A Option 001, E1418A Option 003.

Values of n outside these ranges will generate an error.

The

<

value

>

parameter is a floating point 32-bit number in the range of -16 Volts

to +16 Volts.

MIN

sets -16 Volts.

MAX

sets +16 Volts.

DEF

sets 0 Volts.

This command sets channel n to output the voltage specified following a valid
trigger condition. Following this command, the trigger system must be enabled
using an

INITiate

command. See the

TRIGger

subsystem discussion, page 54, for

additional information.

The channel output relay must be closed for the voltage to appear at the output
terminals.

If the module is in a wait-for-trigger state (

INIT

has been received) when this

command is received, an error is generated.

If the channel is in the current mode when this command is received, an error is
generated.

Related Commands:

SOURce

n

:VOLTage[:LEVel]:TRIGgered[:AMPLitude]?

,

SOURce

n

:FUNCtion:MODE VOLTage

SOURce

Chapter 3

HP E1418A SCPI Command Reference 93

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com