SOURce
n
:CURRent[:LEVel][:IMMediate][:AMPLitude]
SOURce
n
:CURRent[:LEVel][:IMMediate][:AMPLitude] <
value
| MIN | MAX | DEF>
immediately outputs the current set in value at channel n when the output relay is closed.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
SOURce
n
keyword
1 to 8 or 1 to 16
1
<
value
>
numeric (float 32) or
discrete
-0.02 Amps to +0.02 Amps or
MIN | MAX | DEF
0
Comments
•
In the keyword,
SOURce
n, the valid range of n is:
1 to 8 for
E1418A, E1418A Option 002
1 to 16 for
E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
•
The
<
value
>
parameter is a floating point 32-bit number in the range of -0.02 A
to +0.02 A.
MIN
sets -0.02 Amps.
MAX
sets +0.02 Amps.
DEF
sets 0 Amps.
•
This command immediately programs the specified current on channel n. The
output state of the channel must be enabled for the current to appear at the output
terminals.
•
If the channel is in the voltage mode when this command is received, an error is
generated.
•
Related Commands:
SOURce
n
:CURRent[:LEVel][:IMMediate][:AMPLitude]?
,
SOURce
n
:FUNCtion:MODE CURRent
,
OUTPut
n
[:STATe]
SOURce
Chapter 3
HP E1418A SCPI Command Reference 85