A returned value of 5 (calibration aborted) is given if either:

a. The command is received when the module is in voltage calibration or

resistance calibration mode, or

b. The calibration in process has entered more than 50 values without a

successful calibration.

If the internal calibration resistor is not calibrated, the channel is not marked as
calibrated following the completion of calibration (see the

CALibration:CONDition?

query). The new calibration constants are used, however.

When the channel current calibration is completed, the new calibration constants
are used by the module, but not stored in non-volatile memory.

If a calibration is interrupted before completion (that is, by another command
being received), no constants will be changed.

Refer to Appendix D, “ Voltage/Current Output Adjustment” , starting on page 154,
for additional information about this command.

Related Commands:

CALibration:STORe

,

CALibration

n

:CONFigure:CURRent?

CALibration

n

:VALue:VOLTage?

CALibration

n

:VALue:VOLTage? <

value

>

enters a measured calibration value and

returns a value indicating the present state of the voltage calibration process.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

CALibration

n

keyword

1 to 8 or 1 to 16

1

<

value

>

numeric (float 32)

0 to 19.2 V

none

Returns: unsigned 16-bit integer (uint16).

Comments

In the keyword,

CALibration

n, the valid range of n is:

1 to 8 for

E1418A, E1418A Option 002

1 to 16 for

E1418A Option 001, E1418A Option 003.

Values of n outside these ranges will generate an error.

This command is used together with the

CALibration

n

:CONFigure:VOLTage?

query

to perform voltage mode calibration. Following the

CAL

n

:CONF:VOLT?

query, this

command is used repetitively until the return value indicates that the calibration is
complete.

The <value> parameter is the measured output voltage. A minimum of
9 measurements must be taken to calibrate a voltage output channel
(a maximum of 50 measurements may be required).

CALibration

74 HP E1418A SCPI Command Reference