Setting the output mode – HP E1418A User Manual
Page 53
Setting the Output Mode
The output mode can be set using commands in the
SOURce
subsystem.
SOURce
n
:FUNCtion:MODE VOLTage
SOURce
n
:FUNCtion:MODE CURRent
The mode can also be set on the terminal module with the P/J jumpers and
V/I jumpers (see page 31). If the mode is set to Jumper selectable with the
P/J jumper on the terminal module, the mode cannot be changed with the
SCPI commands and the output mode is set by the V/I jumper (also on the
terminal module, see page 31).
For example, if on the terminal module, channel 1 P/J jumper is set to
Jumper Selectable and the V/I jumper is set to current mode, the following
command will each generate an error:
SOURce1:FUNCtion:MODE VOLT
The error returned is:
+3012
Channel
cannot
be
programmed
to
this
mode
,
check
jumpers
Appendix C, starting on page 150 contains a full list of the errors that may
be generated by the SCPI commands for the HP E1418A.
52 Programming Examples
Chapter 2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com