beautypg.com

HP E1418A User Manual

Page 161

background image

Current output adjustment uses two of the

CALibration

subsystem queries.

Current adjustment requires voltage measurements.

CALibration

n

:CONFigure:CURRent?

CALibration

n

:VALue:CURRent? <

value

>

Once the calibration resistance value is set as described earlier, the general
procedure for each channel current output adjustment is:

1. Make the connections shown in Figure D-5 or D-6.
2. Send the

CAL

n

:CONF:CURR?

query to the module and read the

response until a 1 is returned (typically the first query response).

3. Read the voltage output on the DMM.
4. Send the DMM value read with the

CAL

n

:VAL:CURR? <

value

>

query.

5. Enter the query response.
6. Repeat steps 3, 4, and 5 until the

CAL

n

:VAL:CURR? <

value

>

query

returns a 0.

7. Repeat steps 1 through 6 for each channel.

The first query,

CAL

n

:CONF:CURR?

, sets the channel to the calibration

mode and returns a 1 when the channel is ready for calibration. When a 1 is
returned, the channel output voltage is set to the first adjustment point.

Read the output voltage on the DMM and return the value obtained using
the

CAL

n

:VAL:CURR? <

value

>

query. This query returns an integer

indicating the calibration state of the channel. A value other than 0 or 2
from this query indicates additional values are needed.

Figure D-4. Resistance Calibration Connections (CAL)

Appendix D

Voltage/Current Output Adjustment 160

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com