General characteristics, Power requirements, Cooling requirements – HP E1418A User Manual
Page 113

General
Characteristics
Settling Time:
300
µ
S (+full scale to –full scale step, single channel, to accuracy listed
previously)
Isolation (for isolated channels):
42 Vdc/42 V peak (channel-to-chassis or channel-to-channel)
Synchronization:
Software commands, External TTL trigger, or VXIbus TTL Trigger lines
Single channel or multiple channel
VXI Specification:
Device Type: A16 or A24, D16, slave only, register-based
Power
Requirements
Current Required (Amps)
Configuration
+ 5 V
DC
+ 5 V
Dynamic
+ 12 V
DC
+ 12 V
Dynamic
+ 24 V
DC
+ 24 V
Dynamic
– 24 V
DC
– 24 V
Dynamic
8 Channel
Non-Isolated
0.70
0.01
0.04
0.01
0.44
0.01
0.44
0.01
8 Channel
Isolated
0.75
0.01
0.04
0.01
0.88
0.01
0.00
0.00
16 Channel
Non-Isolated
1.25
0.01
0.04
0.01
0.88
0.01
0.88
0.01
16 Channel
Isolated
1.40
0.01
0.04
0.01
1.60
0.01
0.00
0.00
Cooling
Requirements
For a 10
°
C rise
Air Flow
∆
P
Avg
Watts/slot
8 Channel
configurations
2.0
liters/second
0.10 mm H
2
O
25.4
16 Channel
configurations
3.9
liters/second
0.18 mm H
2
O
49.0
112 HP E1418A Specifications
Appendix A
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com