Source n :current[:level][:immediate][:amplitude, Source – HP E1418A User Manual
Page 87
SOURce
n
:CURRent[:LEVel][:IMMediate][:AMPLitude]?
SOURce
n
:CURRent[:LEVel][:IMMediate][:AMPLitude]? [
queries the last output value written by
SOURce
n
:CURRent[:LEVel][:IMMediate][:AMPLitude]
.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
SOURce
n
keyword
1 to 8 or 1 to 16
1
discrete (optional)
MIN | MAX | DEF
none
Returns: 32-bit floating point number (float 32).
Comments
•
In the keyword,
SOURce
n, the valid range of n is:
1 to 8 for
E1418A, E1418A Option 002
1 to 16 for
E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
•
If no parameter is given, this query returns the source current set on channel n by
the last
SOURce
n
:CURRent[:LEVel][:IMMediate][:AMPLitude]
command. The
value returned is the value last programmed on that channel, not an actual
measurement of the hardware output value.
•
If a parameter is given,
MIN
returns -0.02 Amps.
MAX
returns +0.02 Amps.
DEF
returns 0 Amps.
•
This command generates an error if channel n is in the voltage output mode.
•
Related Commands:
SOURce
n
:CURRent[:LEVel][:IMMediate][:AMPLitude]
,
SOURce
n
:FUNCtion:MODE CURRent
SOURce
86 HP E1418A SCPI Command Reference
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com