beautypg.com

Source n :current[:level][:immediate][:amplitude, Source – HP E1418A User Manual

Page 87

background image

SOURce

n

:CURRent[:LEVel][:IMMediate][:AMPLitude]?

SOURce

n

:CURRent[:LEVel][:IMMediate][:AMPLitude]? [

queries the last output value written by

SOURce

n

:CURRent[:LEVel][:IMMediate][:AMPLitude]

.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

SOURce

n

keyword

1 to 8 or 1 to 16

1

discrete (optional)

MIN | MAX | DEF

none

Returns: 32-bit floating point number (float 32).

Comments

In the keyword,

SOURce

n, the valid range of n is:

1 to 8 for

E1418A, E1418A Option 002

1 to 16 for

E1418A Option 001, E1418A Option 003.

Values of n outside these ranges will generate an error.

If no parameter is given, this query returns the source current set on channel n by
the last

SOURce

n

:CURRent[:LEVel][:IMMediate][:AMPLitude]

command. The

value returned is the value last programmed on that channel, not an actual
measurement of the hardware output value.

If a parameter is given,

MIN

returns -0.02 Amps.

MAX

returns +0.02 Amps.

DEF

returns 0 Amps.

This command generates an error if channel n is in the voltage output mode.

Related Commands:

SOURce

n

:CURRent[:LEVel][:IMMediate][:AMPLitude]

,

SOURce

n

:FUNCtion:MODE CURRent

SOURce

86 HP E1418A SCPI Command Reference

Chapter 3