Trigger:source, Trigger – HP E1418A User Manual
Page 106

TRIGger:SOURce
TRIGger:SOURce
sets the trigger source for the module.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
<
source
>
discrete
TTLT0, TTLT1, TTLT2, TTLT3, TTLT4,
TTLT5, TTLT6, TTLT7, BUS, HOLD,
IMM, or EXT
none
Comments
•
The
<
source
>
parameter defines the following trigger conditions:
<
source
>
Trigger
TTLT0–TTLT7
Sets the trigger source to one of the VXIbus backplane trigger lines.
BUS or HOLD
Sets software trigger. A *TRG, TRIG:IMM, or GET will trigger the output.
IMMediate
Places the trigger system in the wait-for-trigger state and then triggers the output.
EXTernal
Sets the trigger source to the EXTernal trigger line on the terminal module.
•
If this command is received when the module is in the wait-for-trigger state, an
error is generated.
•
This command set the source for the trigger in to the module. The module can also
output a trigger on the VXIbus TTLTrg lines using the
OUTPut:TTLTrg
n command.
•
Triggering is described in more detail on page 54.
•
Related Commands:
TRIGger:SOURce?
•
*RST Condition: Trigger source set to
HOLD
.
TRIGger:SOURce?
TRIGger:SOURce?
queries the trigger source for the module.
Parameters
Returns: string.
Comments
•
The query returns one of the following unquoted strings:
String
Trigger Source
TTLT
n
VXIbus backplane trigger line
n
BUS
Software trigger
HOLD
Trigger hold (reset state)
IMM
Immediate trigger
EXT
External trigger
TRIGger
Chapter 3
HP E1418A SCPI Command Reference 105
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com