Controlling the output relay – HP E1418A User Manual
Page 54
Controlling the Output Relay
There are three programming methods to enable the output of a channel
(close the output relay).
•
The
APPLy
n
:VOLTage <
value
>
or
APPLy
n
:CURRent <
value
>
command both program an output value and close the output relay
on a single channel.
•
The
OUTPut
n
:STATe ON
command closes the output relay for a
single channel. The last programmed mode and output value remains
in effect.
•
The
DIAGnostic:OUTPut:ALL:STATe ON
command closes all output
relays. The order of relay closure is not guaranteed.
There are three programming methods to disable the output of a channel
(open the output relay).
•
Send the
OUTPut
n
:STATe OFF
command to open a single channel
output relay without changing the programmed mode or output value
for that channel or affecting any other channels.
•
Send
*RST
to program all channels to voltage mode, 0 volts output,
and open all output relays.
*RST
will not change the output mode if
the P/J jumper is set to Jumper Selectable.
•
Send the
DIAGnostic:OUTPut:ALL OFF
command to open all output
relays without changing the programmed mode or output value.
The output relay will also be opened and then closed by a change of output
mode using the
APPLy
commands.
•
An
APPLy
n
:VOLTage <
value
>
followed by an
APPLy
n
:CURRent <
value
>
command (or vice versa) will open and
then close the output relay.
All output relays will be opened, and left open, by the following commands.
•
The
SOURce
n
:FUNCtion:MODE VOLTage
or
SOURce
n
:FUNCtion:MODE CURRent
commands set output mode
for a single channel and opens the output relay for that channel.
Other channels are not affected.
•
The
*RST
command performs a reset and leaves the output relays
open on all channels (the command also programs all channels to
voltage mode, 0 volts output unless the P/J jumper is set to Jumper
Selectable).
•
The
*TST?
command performs a self-test and leaves the output
relays open on all channels. Programmed output modes are restored
following the self-test.
Chapter 2
Programming Examples 53