DIAGnostic:OUTPut:ALL[:STATe]

DIAGnostic:OUTPut:ALL[:STATe] <

state

>

opens or closes all channel output

relays at once.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

<

state

>

boolean

OFF | ON | 0 | 1

0

Comments

This command performs the same function as the

OUTPut

n

[:STATe]

command,

but enables or disables all outputs simultaneously.

Related Commands:

OUTPut

n

[:STATe]

DIAGnostic:SOURce

n

:FUNCtion:MODE

DIAGnostic:SOURce

n

:FUNCtion:MODE <

mode

>

changes the output mode of the

channel without disabling the channel output.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

SOURce

n

keyword

1 to 8 or 1 to 16

1

<

mode

>

discrete

CURRent or VOLTage

none

Comments

In the keyword,

SOURce

n, the valid range of n is:

1 to 8 for

E1418A, E1418A Option 002

1 to 16 for

E1418A Option 001, E1418A Option 003.

Values of n outside these ranges will generate an error.

Use this command with caution. The output voltage or current is not
predictable during the mode change and transients will occur.

This command performs the same function as the

SOURce

n

:FUNCtion:MODE

command, but does not disable the output during the mode change. This
command can reduce the settling time required to disable and enable the outputs.

Following a mode change with this command, the appropriate calibration constants
are used, the channel mode is changed, and the output is set to either 0 V or 0 A.

This command will generate an error if the trigger system is waiting for a trigger
event when this command is received.

This command is affected by the state of the P/J jumper on the terminal module.
If the P/J jumper is not set to Program Selectable, this command will generate an
error. This jumper is described on page 31.

Related Commands:

SOURce

n

:FUNCtion:MODE

DIAGnostic

78 HP E1418A SCPI Command Reference