CALibration:CONFigure:RESistance?
CALibration:CONFigure:RESistance?
configures the module to perform
calibration of the internal calibration resistor and returns a
0
or
1
indicating the
readiness of the module.
Parameters
Returns: signed 16-bit integer (int16).
1
= ready,
0
= not ready.
Comments
•
The internal calibration resistor should be calibrated before calibrating the
channels in current mode.
•
This query configures the module for a 4-wire resistance measurement at the CAL
output terminals when the calibration point is set to
CALB
(following a
*RST
or
DIAGnostic:CALibration:OUTPut[:MODE] CALBus
command).
•
This query may also be used to set an external calibration resistor value when
calibrating at each channel output.
•
This query is used together with the
CALibration:VALue:RESistance
command to
calibrate the current calibration resistor.
•
Refer to Appendix D, “ Voltage/Current Output Adjustment” , starting on page 154,
for additional information about this command.
•
Related Commands:
CALibration:VALue:RESistance
CALibration:RESet
CALibration:RESet
reloads, for module use, the calibration constants stored in
non-volatile memory.
Parameters
•
None.
Comments
•
This command, with respect to the calibration constants, performs the same action
as power-on.
•
At power-on, or following this command, calibration constants stored in
non-volatile memory are loaded into the module where they are used during
output operations.
•
If the trigger system is waiting for a trigger when this command is received,
an error is generated.
•
This command opens all output relays.
•
Related Commands:
CALibration:STORe
CALibration
68 HP E1418A SCPI Command Reference