Changing configuration – HP E1418A User Manual
Page 168
![background image](/manuals/90248/168/background.png)
1st Value Returned
Indicates an 8-channel or 16-channel configuration.
Bit
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Use
Not Used
Channels
Channels can have one of two values; all 1’s (decimal 7) for an 8-channel
configuration, and all 0’s (decimal 0) for a 16-channel configuration.
2nd Value Returned
Indicates the type of Terminal Module (if any) installed. See page 77 for
additional information.
3rd Value Returned
Indicates isolated or non-isolated plug-on modules for each channel.
Bit
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Use Ch16 Ch15 Ch14 Ch13 Ch12 Ch11 Ch10 Ch9
Ch8
Ch7
Ch6
Ch5
Ch4
Ch3
Ch2
Ch1
When a bit is set to a 0 an isolated plug-on is present. A bit set to a 1
indicates a non-isolated plug-on. For 8-channel devices, channels 9 through
16 bits are set to 1.
4th Value Returned
Indicates the state of the V/I jumper on the terminal module as either
current or voltage when the P/J jumper is in the jumper selectable position.
See page 77 for additional information.
5th Value Returned
Indicates the state of the channel output relay. See page 77 for additional
information.
6th Value Returned
Indicates the state of the P/J jumper on the terminal module. See page 77
for additional information.
Changing
Configuration
Configuration changes are made using Hewlett-Packard add-on kits. These
procedures support the following add-on kits:
Kit Part Number
Description
Use
HP E1523A
1-Channel Isolation
Plug-On Module
Change one or more channels
from Non-Isolated to Isolated
Output.
HP E1524A
8-Channel
Non-Isolated
Expansion Kit
Add 8 Channels of Non-Isolated
Output.
HP E1525A
8-Channel Isolated
Expansion Kit
Add 8 Channels of Isolated
Output.
167 Configuration and Disassembly
Appendix E
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com