beautypg.com

Channel current gain calibration registers – HP E1418A User Manual

Page 144

background image

Channel Current

Gain Calibration

Registers

The Channel Current Gain Calibration Registers store current gain
calibration constants in non-volatile memory. There are 16 registers,
one for each output channel. The register at base + 160

16

stores channel 1,

base + 162

16

stores channel 2, up to base 17E

16

to store channel 16. This

register is non-volatile.

Address

Base + 160

16

through

Base + 17E

16

Bit

15

14

13

12

11

10

09

08

07

06

05

04

03

02

01

00

msb

lsb

WRITE

Unused

Current Gain Calibration Constant

Current Gain Calibration Constant: Writing a 16-bit value to this field
sets a new constant used to calibrate the channel output. Values are in the
range:

000

16

minimum gain

800

16

nominal gain

FFF

16

maximum gain

Notes

Registers 00100

16

through 001BE

16

are located in non-volatile (FLASH)

memory. Writing to this memory requires a complicated sequence of
register accesses not documented here. It is recommended that calibration
be performed using the SCPI driver for the module, not register access.

Non-volatile memory has finite life, based upon the number of writes.
Use care when writing to this memory.

The contents of this register are loaded into volatile memory at power-on
and not used on an on-going basis.

Appendix B

HP E1418A Register-Based Programming 143