beautypg.com

Peak detect, High resolution, Peak detect -4 – RIGOL MSO/DS2000A Series User Manual

Page 80: High resolution -4

background image

RIGOL

Chapter 4 To Set the Sample System

4-4

MSO2000A/DS2000A User’s Guide

Peak Detect


In this mode, the oscilloscope acquires the maximum and minimum values of the
signal within the sample interval to get the envelope of the signal or the narrow pulse
of the signal that might be lost. In this mode, signal confusion can be prevented, but
the noise displayed would be larger.

High Resolution


This mode uses a kind of ultra-sample technique to average the neighboring points
of the sample waveform to reduce the random noise on the input signal and generate
much smoother waveforms on the screen. This is generally used when the sample
rate of the digital converter is higher than the storage rate of the acquisition memory.

Note:

“Average” and “High Res” modes use different averaging methods. The

former uses “Multi-sample Average”

and the latter uses “Single Sample Average”.