Horizontal reference, Horizontal reference -9 – RIGOL MSO/DS2000A Series User Manual

Page 75

background image

Chapter 3 To Set the Horizontal System

RIGOL

MSO2000A/DS2000A User’s Guide

3-9

Horizontal Reference


Horizontal reference is the reference position according to which the screen
waveform is expanded or compressed horizontally when adjusting HORIZONTAL

SCALE. In Y-T mode and the delayed sweep is disabled, press MENUHorRef

in the horizontal control area (HORIZONTAL) at the front panel to select the desired
reference mode and the default is “Center”.

Note:

This function is not available in X-Y mode and Roll mode as well as in Y-T mode

when the delayed sweep is enabled.

1. Center

When changing the horizontal time base, the waveform is expanded or
compressed horizontally around the center of the screen.

2. Trig Pos

When changing the horizontal time base, the waveform is expanded or
compressed horizontally around the trigger point.

3. User

When changing the horizontal time base, the waveform is expanded or
compressed horizontally around the user-defined reference position.

After this mode is selected, press RefPos and rotate

to set the user-defined

reference position. The range is from -350 (the rightmost of the screen) to 350
(the leftmost of the screen) and the default is 0 (the center of the screen). Press

down to restore the reference position to 0.