beautypg.com

To set test results output type, To set test results output type -3 – RIGOL MSO/DS2000A Series User Manual

Page 205

background image

Chapter 10 Pass/Fail Test

RIGOL

MSO2000A/DS2000A User’s Guide

10-3

horizontal and vertical adjustment ranges are 0.02 div to 4.0 div and 0.04 div to 5.12
div respectively.

Figure 10-1 Creat Mask

To Set Test Results Output Type


You can set the output type of the test results referring to the following methods.

1. The Statistic Message

Press Message to select “ON” or “OFF”. When “ON” is selected, the statistic
message of the test results (including the number of the failed frames, the
passed frames and the total frames) will be displayed at the upper right corner
of the screen, as shown in the figure below.


2. Stop on Fail

Press StopOnFail to select “ON” or “OFF”.
 ON: when a failed waveform is detected, the oscilloscope will stop the test

and enter the “STOP” state. At this point, the results of the test remain the
same on the screen (if display is turned on) and only one negative pulse is
output from the [Trigger Out] connector (if enabled) at the rear panel.

 OFF: the oscilloscope will continue with the test even though failed