RIGOL MSO/DS2000A Series User Manual
Page 122
![background image](https://www.manualsdir.com/files/859100/content/doc122.png)
RIGOL
Chapter 5 To Trigger the Oscilloscope
5-36
MSO2000A/DS2000A User’s Guide
Setup Type:
Press SetupType to select the desired setup type.
Setup: set the time that the data stays stable and constant before the clock
edge appears. Press Setup to set the setup time and the range is from 2 ns to 1
s.
Hold: set the time that the data stays stable and constant after the clock edge
appears. Press Hold to set the hold time and the range is from 2 ns to 1 s.
SetupHold: set the time that the data stays stable and constant before and
after the clock edge appears. Press Setup and Hold to set the setup time and
hold time respectively and the range is from 2 ns to 1 s.
Trigger Mode:
Press Sweep to open the trigger mode list and select “Auto”, “Normal” or ”Single”.
For the details, please refer to “Trigger Mode”. The corresponding status light of
the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.
Note:
Noise rejection is grayed out and disabled automatically when the signal
source is set to any channel of D0-D15.
Trigger Level:
Press DataSrc and use TRIGGER LEVEL to modify the trigger level of the data
source channel. Press ClkSrc and use TRIGGER LEVEL to modify the trigger
level of the clock source channel. For the details, please refer to the introduction of