RIGOL

Chapter 10 Pass/Fail Test

10-2

MSO2000A/DS2000A User’s Guide

To Enable Pass/Fail Test


Press UtilityPass/FailEnable to select “ON”. The blue shadow area denotes
the failure area and if the waveform measured at certain time point during the
measurement process passes through the failure area, it will be treated as failed
frame. Press Operate to select “ ” to start testing and select “■” to stop testing.

Note:

The pass/fail test function is not available in the following status.

 In X-Y time base mode
 In ROLL time base mode
 In Y-T time base mode and the horizontal time base is set to 200ms/div or slower,

namely the instrument enters “Slow Sweep” mode

 When the waveform record enabled

After the pass/fail test is enabled, you can set the output type of test results
according to “To Set Test Results Output Type”. You can also select the signal source,
set the test mask range, set the output form of the test results, create mask as well
as save and recall the test mask. For the details, refer to the introductions below.

To Select Source


Press Source to select the channel (CH1 or CH2) to be tested and only channels
enabled can be selected.

Mask Range


Users can define their desired test masks.

Press RangeRange to select “Screen” or “Cursor”. The default is “Screen”. When
“Cursor” is selected, two gray cursor lines appear on the screen. At this point, press
CursorA and CursorB; then, use

to adjust the positions of the two cursor lines

respectively. Or, press CursorAB and use

to adjust the positions of cursor A and

cursor B at the same time.
Press X Mask and Y Mask, rotate

and the mask lines appear on the screen as

shown in the figure below. Press Create to apply the mask currently created. The