RIGOL MSO/DS2000A Series User Manual
Page 303
RIGOL
MSO2000A/DS2000A User’s Guide
18-5
Trigger
Sensitivity
1 div (below 10 mV or noise rejection is enabled)
0.3 div (above 10 mV and noise rejection is disabled)
Edge Trigger
Edge Type
Rising, Falling, Rising/Falling
Pulse Trigger
Pulse
Condition
Positive Pulse Width (greater than, lower than, within specific
interval)
Negative Pulse Width (greater than, lower than, within specific
interval)
Pulse Width
Range
2 ns to 4 s
Runt Trigger
Pulse Width
Condition
None, >, <, <>
Pulse Polarity
Positive, Negative
Pulse Width
Range
2 ns to 4 s
Windows Trigger (Option)
Windows Type Rising, Falling, Rising/Falling
Trigger
Position
Enter, Exit, Time
Windows
Time
16 ns to 4 s
Nth Edge Trigger (Option)
Edge Type
Rising, Falling
Idle Time
16 ns to 4 s
Edge Number 1 to 65535
Slope Trigger
Slope
Condition
Positive Slope (greater than, lower than, within specific interval)
Negative Slope (greater than, lower than, within specific interval)
Time Setting
10 ns to 1 s
Video Trigger (HDTV Option)
Signal
Standard
NTSC, PAL/SECAM, 480P, 576P (Standard)
720P, 1080P and 1080I (Option)
Pattern Trigger
Pattern
H, L, X, Rising, Falling