Chapter 6 MATH and Measurements
RIGOL
MSO2000A/DS2000A User’s Guide
6-21
“Delay 1 →2 ”, “Delay 1 →2 ” and “Delay 1 →2 ”.
Press Measure Setting Type “Delay” and then press SourceA and
SourceB to set the two channel sources (CH1, CH2 or any channel of D0-D15)
of delayed measurement respectively.
Note:
When SourceA is set to any analog channel (or digital channel),
SourceB can only be set to a analog channel (or digital channel).
4. Phase Measurement Setting
Specify the source A and source B in the measurement items “Phase 1 →
2 ”,
“Phase 1 →
2 ”, “Phase 1 →2 ” and “Phase 1 →2 ”.
Press Measure Setting Type “Phase” and then press SourceA and
SourceB to set the two channel sources (CH1, CH2 or any channel of D0-D15)
of phase measurement respectively.
Note:
When SourceA is set to any analog channel (or digital channel),
SourceB can only be set to a analog channel (or digital channel).
5. Threshold Measurement Setting
Specify the upper limit, lower limit and middle value of the auto measurement of
the analog channel. Measurements of all the time, delay and phase parameters
will be influenced by this setting.
Press Measure Setting Type “Threshold” and then:
Press Max and use
to set the maximum value of the measurement.
Reducing the maximum value to the current “Mid” will automatically reduce
“Mid” and “Min” to keep them lower than the maximum value. The default is
90% and the range available is from 7% to 95%.
Press Mid and use
to set the middle value of measurement. The middle
value is limited by the settings of “Max” and “Min”. The default is 50% and
the range available is from 6% to 94%.
Press Min and use
to set the minimum value of the measurement.
Increasing the minimum value to the current “Mid” will automatically
increase “Mid” and “Max” to keep them higher than the minimum value.
The default is 10% and the range available is from 5% to 93%.