Measurement Setting
1. Source Selection
Press Measure Source to select the desired channel for measurement (CH1,
CH2, MATH or any channel of D0-D15). The color of the parameter icons under
MENU at the left of the screen will change with the source selected.
2. Measurement Range
Press Measure Range Range to select “Screen” or “Cursor” for
measurement.
When “Cursor” is selected, two cursor lines appear on the screen. At this point,
press CursorA and CursorB and use
to adjust the position of the two
cursor lines respectively so as to determine the measurement range. Or, press
CursorAB and use
to adjust the positions of cursor A and cursor B at the
same time.
Note:
You can press
continuously to switch among CursorA, CursorB and
CursorAB.
Figure 6-6 “Cursor” Selected for Measurement Range
3. Delayed Measurement Setting
Specify the sourceA and sourceB in the measurement items “Delay 1 →
2 ”,