Analysis Based on Pass/Fail Mask


Press Analyze and select “Pass/Fail”. Then, set the templet used in analysis based
on Pass/Fail mask through the method below.

Press Range to open the following setting menus.

1. Range

Press this softkey to select “Screen” or “Cursor” for the analysis (the default is
“Screen”). When “Cursor” is selected, two gray cursor lines appear on the screen.
At this point, press Cursor A and Cursor B; then, use

to adjust the

positions of the two cursor lines respectively to determine the analysis range. Or,
press CursorAB and use

to adjust the positions of cursor A and cursor B at

the same time.

Note:

You can press down

continuously to switch the current cursor.


2. X Mask

Press this softkey and use

to adjust the horizontal threshold and the range

is from 0.02 div to 4.00 div. During the adjustment, two curves would be
displayed to show the outline of the current mask as shown in the figure below.

Figure 11-4 Analysis Based on Pass/Fail Mask


3. Y Mask

Press this softkey and use

to adjust the vertical threshold and the range is

from 0.04 div to 5.12 div. During the adjustment, two curves would be displayed
to show the outline of the current mask.

Outline of the Current Mask