beautypg.com

RIGOL MSO/DS2000A Series User Manual

Page 152

background image

RIGOL

Chapter 6 MATH and Measurements

6-16

MSO2000A/DS2000A User’s Guide

below.

8. Phase 1 2 : phase difference calculated according to “Delay 1 →2 ” and

the period of source 1, expressed in degree. The calculation formula is as shown
below.

The phase calculation formula:

360

1

Delay

Phase

Period

=

×

°

Wherein,

Phase

represents the pahase measurement result

Delay

represents the delay measurement result

1

Period

represents the period of source 1