Self-calibration, Self-calibration -11 – RIGOL MSO/DS2000A Series User Manual
Page 279
Chapter 15 System Function Setting
RIGOL
MSO2000A/DS2000A User’s Guide
15-11
Self-calibration
The self-calibration program can quickly make the oscilloscope reach the best
working state to get the most precise measurement values. You can perform
self-calibration at any time especially when the change of the environment
temperature is up to or more than 5 ℃. Make sure that the oscilloscope has been
warmed up or operated for more than 30 minutes before the self-calibration.
Disconnect all the input channel connections, and then press Utility Self-Cal and
the self-calibration interface as shown in the figure below is displayed.
Figure 15-4 Self-calibration Interface
Press Start and the oscilloscope will start to execute the self-calibration program.
Press Exit to give up the self-calibration operation at any time and return to the
previous menu.
Note:
Most of the keys are disabled during the self-calibration.