Probe Compensation


When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may cause measurement inaccuracy or error. The probe
compensation procedures are as follows.

1. Perform steps 1, 2, 3 and 4 of Function Inspection.
2. Check the displayed waveforms and compare them with the following figures.

Figure 1-9 Probe Compensation

3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment

hole on the probe until the displayed waveform is as the “Perfectly
compensated” in the figure above.


















Over compensated Perfectly compensated Under compensated