Nth edge trigger (option), Nth edge trigger (option) -18 – RIGOL MSO/DS2000A Series User Manual
Page 104
RIGOL
Chapter 5 To Trigger the Oscilloscope
5-18
MSO2000A/DS2000A User’s Guide
Nth Edge Trigger (Option)
Trigger on the nth edge that appears after the specified idle time. For example, in the
waveform shown in the figure below, the instrument should trigger on the second
rising edge after the specified idle time (the time between two neighbouring rising
edges) and the idle time should be set to P
between the rising edges participate in the counting.
P
M
P Edge Type= Edge Num=2 Figure 5-5 Nth Edge Trigger to select “Nth Edge” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen. For example, . The trigger type is Nth edge trigger; the trigger source is CH1; the trigger level is 1.94 V. Press Source and select CH1, CH2 or any channel of D0-D15. For the details, please refer to the introduction in “Trigger Source”. The current trigger source is displayed at the upper right corner of the screen. Note: Select channel with signal input as trigger source to obtain stable trigger. triggers. : trigger on the rising edge of the input signal when the voltage level meets the specified trigger level.
Trigger Type:
Press Type, rotate
Source Selection:
Edge Type:
Press Slope to select the kind of edge of the input signal on which the oscilloscope