General specifications – RIGOL MSO/DS2000A Series User Manual
Page 309
![background image](https://www.manualsdir.com/files/859100/content/doc309.png)
RIGOL
MSO2000A/DS2000A User’s Guide
18-11
General Specifications
Probe Compensation Output
Output Voltage
[1]
About 3 V, peak-peak
Frequency
[1]
1 kHz
Power
Power Voltage
100 V to 240 V, 45 Hz to 440 Hz
Power
Maximum 50 W
Fuse
2 A, T Degree, 250 V
Environment
Temperature Range
Operating: 0 ℃ to +50 ℃
Non-operating: -40 ℃ to +70 ℃
Cooling Method
Fan cooling
Humidity Range
0℃ to +30℃: ≤95% Relative Humidity
+30℃ to +40℃: ≤75% Relative Humidity
+40℃ to +50℃: ≤45% Relative Humidity
Altitude
Operating: under 3,000 meters
Non-operating: under 15,000 meters
Physical Characteristics
Size
[3]
Width×Height×Depth = 361.6 mm× 179.6 mm×130.8
mm
Weight
[4]
Package Excluded
3.9 kg ± 0.5 kg
Package Included
4.5 kg ± 0.5 kg
Calibration Interval
The recommended calibration interval is one year.
Regulatory Information
Electromagnetic
Compatibility
2004/108/EC
Execution standard EN 61326-1:2006 EN 61326-2-1:2006
Safety
UL 61010-1:2004; CAN/CSA-C22.2 NO. 61010-1-2004;
EN 61010-1:2001; IEC 61010-1:2001
Note
[1]
: Typical.
Note
[2]
: Maximum value. 20ns, single-channel mode, dots display, auto memory depth.
Note
[3]
: Supporting legs and handle folded, knob height included.
Note
[4]
: Standard configuration.