Stacked Volume Test
The Stacked Volume test measures the volume of multiple breaths to determine the
accuracy of the ventilator in accumulating volume.
To access the Stacked Volume test:
1. Press the
MORE/4
key.
2. Use the S and T soft keys to highlight
Stacked Volume Test
.
3. Press the
ENTER
soft key.
4. Press the
MODIFY
soft key.
5. Use the keypad to enter the number of breaths for the test and press the
ENTER
soft
key.
To run the test:
1. Press the
START
or
AUTO
soft key to begin the test. The display changes to show
the progress of the test:
fec067.bmp
The values shown are described below:
Total Volume
– The total volume of all breaths together
Last Volume
– The volume of the last breath taken
Avg. Volume
– The average volume of all breaths taken
Num. Of Breaths
– The total number of breaths taken
Total # Breaths
– The total number of breaths to be taken
Note
If a stacked volume test is started with the
START
soft key, the test
accumulates the requested number of breaths and ends. If the test is started
using the
AUTO
soft key, the test automatically restarts after the requested
number of breaths have accumulated.
2. When the test is complete, press the
STACK
soft key to start a new stacked volume
test or press the
EXIT
soft key to exit.