Leak test – Fluke Biomedical VT Plus HF User Manual
Page 74

VT Plus HF
Operators Manual
3-24
Leak Test
The Analyzer allows the user to test the leak rate of a sealed vessel or test lung. A Leak
test can be done using the high pressure, low pressure, or airway pressure signals to
measure the leak rate of a sealed vessel or test lung.
W
Caution
To avoid damage to the Analyzer, be sure that maximum
pressure ratings are not exceeded when performing a Leak test.
See Specifications for absolute maximum pressure ratings for
each pressure port.
To access the Leak test screen:
1. Press the
MORE/4
key.
2. Use the S and T soft keys to highlight
Leak Test
.
3. Press the
ENTER
soft key
.
The
Leak Test Setup
screen displays.
fec061.bmp
To set test parameters:
1. On the
Leak Test Setup
screen, press the S and T soft keys to highlight
Pressure Channel
.
2. Press the
MODIFY
soft key until the desired pressure channel is selected.
Note
If
Airway
is selected, the airway transducer is connected to the currently
selected flow channel. Both flow channels can be tested; however, the
appropriate channel must be selected using the
RANGE
function in the
Flow Mode
screen.
3. To enter the compliance of the test chamber (test lung) press the S and T soft keys
to highlight
Compliance
.
4. Press the
MODIFY
soft key. Use the keypad to enter the compliance in l/cmH
2
O. Be
sure to use leading zeros so that a number is entered for all five digits.
5. To enter the duration of the test, press the S and T soft keys to highlight
Test
Time
.
6. Press the
MODIFY
soft key.