Other setup functions, Breath detect, Breath detect threshold – Fluke Biomedical VT Plus HF User Manual
Page 115: Gas settings

Remote Operation
Other Setup Functions
4
4-31
Other Setup Functions
VT for Windows allows selection of settings in the
Setup
menu.
Breath Detect
The Analyzer has a built-in breath detection algorithm. The algorithm determines the
breath phase (inspiration, expiration, inspiratory pause, single phase bi-directional) This
algorithm is designed to reject signal noise and other artifact when determining the
phases of a breath.
To change the breath detect mode:
1. From the
Setup
menu, select
Breath Detect
.
2. Click the preferred option.
Breath Detect Threshold
The Analyzer breath detection algorithm uses a flow rate threshold that the flow must
cross to trigger a change in the breath phase. Higher flow rate thresholds yield less
sensitive breath detection. Lower thresholds yield more sensitive breath detection.
To enter a new breath detect threshold:
1. From the
Setup
menu, select
Breath Detect
; then select
Threshold
. The
following dialog box displays:
fec115.bmp
2. Enter a new value and click
OK
.
Gas Settings
Through the
Gas Settings
option, the Analyzer can compensate for the effects of gas
temperature, viscosity, molecular weight, and relative humidity, on flow measurement, as
well as correct flow calculations to one of several standard conditions. For more
information on the gas settings, refer to System Setup.
To change the gas settings:
1. From the
Setup
menu, select
GAS SETTINGS
. The following dialog box appears,
allowing changes to be made: