VT Plus HF
Operators Manual
2-20
Utilities
Note
These settings are required by service technicians in servicing the Analyzer.
To access the utility functions:
1. From the
Setup
screen, highlight
Utilities
and press the
ENTER
soft key. The
Utilities
menu displays:
fec035.bmp
Oxygen Sensor Calibration
Recalibrate the oxygen sensor with 100 % oxygen and dry air at the beginning of any day
in which oxygen concentrations are to be measured.
To calibrate the oxygen sensor:
1. Apply 10 lpm of dry air to the high-flow inlet or exhaust.
2. Use the S and T soft keys to highlight
Oxygen Calibration
.
3. Press the
ENTER
soft key. The
O2 Calibration
screen displays.
fec075.bmp
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)