Fuse replacement – Fluke Biomedical VT Plus HF User Manual
Page 122

VT Plus HF
Operators Manual
5-4
Fuse Replacement
The Analyzer fuses may have to be replaced. The fuse compartment is located on the side
panel of the instrument, as shown in Figure 5-1.
fec122.eps
Figure 5-1. Fuse Compartment
To replace the fuses:
1. Power off Analyzer and unplug the main power cord.
2. Using a small, flat-blade screwdriver or similar tool, carefully pry off the cover to the
fuse compartment. The fuse block assembly is attached to the back of the cover.
3. Pull out the fuse block assembly and carefully lift the old fuses from the fuse slots.
4. Replace the fuses (two 0.5 A Slo-Blo fuses, Fluke part #46026).
XW
Warning
To avoid possible fire, electrical shock, or personal injury, use
only the recommended fuse.
5. Replace the cover/fuse block in the fuse compartment.
6. Plug in the Analyzer.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)