System diagnostics, Linearization – Fluke Biomedical VT Plus HF User Manual
Page 49

Connection and Setup
Utilities
2
2-21
4. Press the
ZERO
soft key.
5. Press
OK
to start 21 % O
2
calibration. Allow the full 2-minute period. Do not press
SKIP
.
6. Verify that the O2 reading is 20.9
±
0.1 %.
7. Press the
CAL
soft key.
8. Apply 10 lpm of dry 100 % oxygen to the high-flow inlet or exhaust.
9. Press
OK
to start 100 % O2 calibration. Allow the full 2-minute period. Do not press
SKIP
.
10. Verify that the O
2
reading is 100
±
0.1 %.
11. Press the
BACK
soft key to return to the
Utilities
menu.
Note
If the oxygen sensor cannot be calibrated, it may need to be replaced. Refer
to Maintenance, Service, and Calibration: Oxygen Sensor Replacement.
System Diagnostics
These functions are for the technical service personnel to use while repairing the
Analyzer. They include
Zero All
,
AD Diagnostics
, and
Signal Diagnostics
.
Linearization
This option is most useful to technical service personnel repairing the Analyzer. There is
no input, but the calibration data for each sensor is stored and displayed, so a checksum
test can be performed to make sure the sensor data is still valid.
To check linearization data:
1. From the
Utilities
menu, highlight
Linearization
. The
Linearizaion Info
screen
displays:
fec036.bmp
2. Press the
BACK
soft key to return to the
Utilities
menu.