Stacked volume test screen – Fluke Biomedical VT Plus HF User Manual
Page 102

VT Plus HF
Operators Manual
4-18
Ending the Leak Test
To end the Leak test:
1. Either click the
STOP
button or let the Leak test end by the time running out. When
the Leak test is over a report is prepared showing the final values for the calculated
parameters.
2. Click the
button to print a final report for the Leak test.
3. Click the
New Test
button to start a new Leak test.
Stacked Volume Test Screen
The Stacked Volume test measures tidal volumes over a given number of breaths and
calculates the average tidal volume.
Setting Stacked Volume Test Parameters
The Stacked Volume test setup screen, shown in Figure 4-13, allows the user to enter the
number of breaths to be averaged.
fec095.bmp
Figure 4-13. Stacked Volume Test: Setup Screen
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)