Plots, Automatically re-scaling plots, Manually re-scaling plots – Fluke Biomedical VT Plus HF User Manual
Page 112

VT Plus HF
Operators Manual
4-28
Plots
Automatically Re-scaling Plots
By default, VT for Windows attempts to re-scale plots so that the entire plot is visible.
When a re-scale occurs, the high- and low-ranges on the plots are determined by the min
and max statistics for the signal shown on the plot. The monitor plots and the single plot
are re-scaled every time they reach the end of the plot. The loop plots are re-scaled on
every breath.
Automatic re-scaling may be turned off by unchecking the box next to the plot name in
the
Plots, Scaling, Auto-Scale
menu, as shown below.
fec110.bmp
Auto-scaling is automatically turned off for a plot if the plot is manually re-scaled. Auto-
scaling can also be turned off by right clicking on a plot and unchecking the
Auto-Scale
option.
Selecting the
Plots, Scaling, Auto-Scale
function, or clicking the
Re-Scale Now
button on the tool bar forces an automatic re-scale to occur immediately.
Manually Re-scaling Plots
There are two ways to specify a specific range to view on a plot:
•
Select
Scaling
and
Manual Scale
from the Plots menu bar, as shown below:
fec111.bmp
Then click the preferred plot. The following dialog box appears: