Gas settings, Gas type – Fluke Biomedical VT Plus HF User Manual
Page 36

VT Plus HF
Operators Manual
2-8
Note
The option displayed when exiting the settings screen is saved and used for
future measurements.
The following is a description of each setting.
Gas Settings
This setting tells the Analyzer what type of gas is flowing through the high- and low-flow
ports. The Analyzer uses this information to improve the accuracy of the flow
calculations.
Note
Failure to enter the correct value causes the flow and volume calculations
to be in error.
To modify the gas settings:
1. From the
Setup
menu, highlight
Settings
and press the
ENTER
soft key.
2. Highlight
Gas Settings
and press the
MODIFY
soft key.
Depending on the option, a value can be pre-selected or input from the keypad. The
following is a description of each of the options.
Gas Type
The possible values for
Gas Type
are shown in Table 2-1.
Table 2-1. Gas Type Values
Value Description
Air
Standard room air
N2
100 % Nitrogen
N2O
100 % Nitrous Oxide
CO2
100 % Carbon Dioxide
O2
100 % Oxygen
Heliox
30 % Oxygen and 70 % Helium blend
N2O bal O2
Measured Oxygen balance Nitrous Oxide
He bal O2
Measured Oxygen balance Oxygen Helium
N2 bal O2
Measured Oxygen balance Nitrogen
User Defined User selects custom gas mixture. This option requires the VT for Wondows
PC software.
Note
For
balance
gas settings, the oxygen concentration is measured using the
Analyzer oxygen sensor located on the high-flow port. If the low-flow port
is to be used, allow gas to flow through both high- and low-flow ports so
that the oxygen can be measured.
To modify the gas type: