Selecting breath parameters – Fluke Biomedical VT Plus HF User Manual
Page 60

VT Plus HF
Operators Manual
3-10
Selecting Breath Parameters
The Analyzer calculates 17 breath parameters for every breath:
fec048.bmp
The user can select up to four of these parameters to be displayed with each of the
screens.
To select parameters for display on a specific screen:
1. Select the screen for which the parameters are to be added. For example, to select
parameters for addition to the
Flow
screen, press the
FLOW/0
key.
2. Press the
PARAMETERS/9
key to display all possible parameters. An
X
in the
small box adjacent to a parameter name indicates that the parameter is to be added.
An empty box indicates that the parameter is not to be added.
3. Use the soft keys labeled Sand T to highlight parameters.
4. Press the
ENTER
soft key to select (add) or deselect (not add) a parameter. Only
four parameters can be viewed on a screen; therefore, an existing parameter must be
deselected from the list before a new parameter can be selected.
5. Repeat steps 3 and 4 above for each additional parameter to be changed. Selecting the
Set all screens to the same parameters
option in the parameters screen
displays the currently selected parameters on all screens. If this option is deselected
after having been in effect, previous parameter selections for each screen must be
reselected.
6. Press the
EXIT
soft key to return to the previous screen.
Note
Parameter selections are retained even after the Analyzer has been
powered off.