Unidirectional flow mode – Fluke Biomedical VT Plus HF User Manual
Page 33

Connection and Setup
Connecting the Analyzer
2
2-5
Unidirectional Flow Mode
Alternatively, the Analyzer can be used to measure only inspiratory or expiratory gas
flow, as described below.
Inspiratory Testing
1. Connect the inspiratory hose to the flow port on the right of the Analyzer.
2. Connect the flow exhaust port on the left to the test lung using standard breathing
hoses.
O
u
tlet
Inlet
V
entilator
Inspiratory Hose
Expiratory Hose
TEST
LU
N
G
V
T PLUS HF
fec008.eps
Figure 2-2. Connecting the Analyzer to the Inspiratory Limb of the Breathing Circuit
3. Ensure that the breath detection mode is set to Unidirectional Inspiratory.
a. Go into
Setup
b. Choose
Settings
c. Choose
Breath Detect
and modify as needed
Expiratory Testing
1. Connect the expiratory hose to the flow port on the right side of the Analyzer.
2. Connect the flow exhaust port on the left to the ventilator using standard breathing
hoses.
Inlet
O
u
tlet
V
entilator
Inspiratory Hose
Expiratory Hose
TEST
LU
N
G
V
T PLUS HF
fec009.eps
Figure 2-3. Connecting the Analyzer to the Expiratory Limb of the Breathing Circuit