Title bar, Data plots, Breath parameters – Fluke Biomedical VT Plus HF User Manual
Page 57

Operation
Screen Objects
3
3-7
Title Bar
Each screen includes a title bar that shows the screen or mode name on the left. A number
indicating the screen number is shown in a small box on the right of the title bar:
fec039.bmp
9
Indicates the screen number
Data Plots
Except the
FULL
screen, each screen includes a data plot that shows a measured signal
(flow, pressure volume, or O
2
) vs. time:
fec045.bmp
New data is added to the plot starting on the left and progressing to the right. When the
plot has progressed to the right edge, the plot restarts on the left side of the plot. One line
of data from the previous plot is removed ahead of the new plot.
When a new screen is selected, the plot is displayed using a default plot range. Pressing
the
CLEAR
soft key restarts the plot and clears all previously plotted data and calculated
statistics. Pressing the
RESCALE
key changes the window of the data plot to a more
appropriate range based on the minimum and maximum values in the Statistics area.
Note
After pressing
CLEAR
, allow one full cycle to pass before pressing
RESCALE
.
Breath Parameters
Every screen can display up to four breath parameters:
fec047.bmp
These parameters are displayed only when breath detection is enabled. See Special
Functions for instructions on how to enable breath detection. Parameters are updated
following every breath. Units of measure for breath parameters are fixed. See
Introduction: Specifications for a listing of breath parameters and units. See Selecting
Breath Parameters for instructions on selecting which parameters are displayed for each
of the modes.