Frequency Response. .................... >25 Hz or t10-90 <40 ms, whichever is greater
Sample Rate................................... 100 Hz
Dynamic Resistance....................... <2.00 cmH
2
O @ 60 lpm
Low-flow Dropout............................ 25 lpm
Breath Detect Threshold................. (user settable) 2.0 lpm
Volume Range................................ >±60 l
Notes
•
Tidal Volume Accuracy ±3 % of reading or ±10 ml, whichever is
greater
•
Volume Accuracy tested to 7 liters.
•
Flow accuracy is specified for dry air or oxygen.
•
Do not use with humidified gases.
Low Flow Port
Maximum Flow Rate........................ (absolute value) 50 lpm
Operating Flow Range.................... ±25 lpm
Accuracy......................................... ±1.0 % of range
Resolution....................................... 0.01 lpm flow >1 lpm; 0.001 lpm flow <1 lpm
Frequency Response ..................... . >25 Hz or t10-90 <40 ms, whichever is greater
Sample Rate................................... 100 Hz
Dynamic Resistance....................... <2.5 cmH
2
O @ 5 lpm
Low-flow Dropout............................ 0.01 lpm
Breath Detect Threshold................. (user settable) 0.50 lpm
Volume Range ............................... >±60 l
Note
•
Tidal Volume Accuracy ± 3 % of reading or ±5 ml, whichever is
greater
•
Volume Accuracy tested to 1 liter
•
Flow accuracy is specified for dry air or oxygen.
•
Do not use with humidified gases.
•
Below 0.25 lpm, measurement accuracy is obtained by allowing the
Analyzer to fully warm up, and manually zeroing before reading or
documenting a measurement.
Breath Parameter Accuracy Specifications
Breath parameter accuracy specifications are listed in Table 1-7.
Note
If a choice of readings is indicated by an “or”, choose whichever is
greater; for example, 3 % or 250 ml.