Table 3-1. Parameters Calculated by the Analyzer
Parameter Abbreviation
How
Calculated
Inspiratory Time
In Time
Time of the inspiratory period including the
inspiratory hold time
Expiratory Time
Ex Time
Time of the expiratory period including the
expiratory hold time
Inspiratory Hold Time
In Hold
Time of the End Inspiratory State
Expiratory Hold Time
Ex Hold
Time of the End Expiratory State
Inspiratory to Expiratory
Time
Ratio I:E
Ratio of the inspiratory time to the Expiratory
Time
Inspiratory Hold
Pressure
IPP
Average pressure during the End Inspiratory
State
Positive End-Expiratory
Pressure
PEEP
Average pressure during the End Expiratory
State
Mean Airway Pressure
MAP
Mean pressure over the entire breath
Peak Inspiratory
Pressure
PIP
Maximum pressure during the Inspiratory Time
Tidal Volume
Tidal Vol
Total volume accumulated over the Inspiratory
Time
Minute Volume
Minute Vol
Expiratory Tidal Volume times the respiratory
rate in breaths per minute; reported as an eight-
breath average
Peak Inspiratory Flow
Rate
PIF
Maximum of the absolute value of the Flow Rate
measured during the Inspiratory Time
Peak Expiratory Flow
Rate
PEF
Maximum of the absolute value of the Flow Rate
measured during the Expiratory Time
Compliance
N/A
Expiratory Tidal Volume divided by the difference
of the Inspiratory Hold Pressure and PEEP