This setting is used to adjust the transition of signals on the flow meters. The filter has
four levels:
Soft
,
Medium
,
Hard
, and
None
.
Note
When filtering is on, full test parameters are invalid.
To change the filtering level:
1. Highlight
Filtering
.
2. Press the
MODIFY
soft key to toggle to the desired value.
3. Press the
ENTER
soft key to save the change or the
ESCAPE
soft key to cancel the
change.
4. Press the
BACK
soft key to return to the
Settings
menu.
Display Calibration Date on Startup Screen
Use this option to control whether the calibration due date appears on the startup screen.
To select whether the calibration due date is shown on the startup screen:
1. Highlight
Display Cal Date
.
2. Press the
MODIFY
soft key to toggle between
YES
and
NO
.
3. Press the
ENTER
soft key to save the change or the
ESCAPE
soft key to cancel the
change.
4. Press the
BACK
soft key to return to the
Settings
menu.
Serial Mode
Use this option to control what data is available through the Analyzer serial port. The
possible options are shown in Table 2-3.
Table 2-3. Serial Port Options
Option Description
Comp Ctrl
Computer Control – specialized serial communications used by the VT for
Windows PC Software; contact your sales representative for the protocol if
needed.
RT200
A serial mode that emulates the Allied RT200 serial communications
None
No serial communications