Troubleshooting
C
C-5
Table C-1. Troubleshooting (cont.)
Problem
Possible Cause
Corrective Action
Analyzer does not print.
Wrong printer type selected
Select correct printer type on the
Analyzer.
Printer not set up for selected
printer language
Review printer's user manual to
determine how to set up printer
to work with the selected
Analyzer printer mode
Printer not connected
Connect printer.
Analyzer takes a long time to
establish communications with
PC software.
This is normal.
If the Analyzer does not
eventually communicate with the
PC software, ensure the cable is
correctly connected and that the
PC hardware can support the
communication port selected.
Make sure no other device is
trying to use the serial port on
your PC. Close any software that
may be using the serial port.
Make sure that Windows
recognizes the serial port in the
Control Panel. Restart VT for
Windows.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)