Fluke Biomedical VT Plus HF User Manual
Page 103

Remote Operation
Main Screens and Tests
4
4-19
Running the Stacked Volume Test
To run the stacked volume test:
1. Click the
START
button to begin testing. The screen changes, as shown in Figure
4-14:
fec096.bmp
Figure 4-14. Stacked Volume Screen: Testing Underway
When the stacked volume test begins, the Analyzer switches into Inspiratory Only
breath-detect mode. At the end of the test, the breath-detect mode is restored to its
previous state. The stacked volume test reports five values to the user; they are
updated every time a breath is detected:
Total Volume
–the total of the tidal volumes for the breaths tested
Last Volume
–the tidal volume for the last breath
Average Volume
– the average tidal volume over the breaths that have been tested and
calculated
Number of Breaths
– the number entered at the start of the test; it does not change during
a test.
Breath Count
–the current number of breaths included in the test; if tidal volume can't be
calculated for a breath, that breath and volume are not included in the test.