Volume screen – Fluke Biomedical VT Plus HF User Manual
Page 66

VT Plus HF
Operators Manual
3-16
Volume Screen
The Volume screen displays the signal and statistics for the volume signal.
fec053.bmp
The volume signal is the integral of the flow signal measured in either the high or low
flow ports. The flow port that is integrated is selected by selecting the high- or low-flow
range from within the flow screen.
Flow Zero
Press the
FLOW ZERO
soft key to zero the flow reading.
Units Press
the
UNITS
soft key to change the units of measure for the volume
screen. Possible units of measure for this screen include:
l
Liters
ml
Milliliters
CF
Cubic
Feet
Re-scale Press
the
RESCALE
soft key to change the scale on the plot of data to a
more appropriate window of values.
Note
RESCALE
uses the current min/max values to
determine plot limits.
Clear Press
the
CLEAR
soft key to reset the statistics and re-start the plot.
The user can select the breath parameters displayed on this screen by following the
instructions under Selecting Parameters.
Note
The volume signal is integrated from either the high- or low-flow signals.
To select which flow signal is used to calculate volume:
1. Press the
FLOW/0
key.
2. Press the
RANGE
soft key until the screen corresponding to the desired flow signal
(high or low) displays.