Flowserve 3400IQ Digital Positioner User Manual
Page 33

33
Logix 3400IQ Digital Positioner FCD LGENIM3402-00 – 0/07
flowserve.com
Linear vs. Rotary	The	positioner	has	two	configuration	settings:	Linear	and	Rotary.	In	order	to	
get	better	resolution,	stem	position	sensor	gains	are	adjusted	based	on	the	angle	of	rotation	of	the	
linkage.	The	linear	setting	allows	for	linkage	rotation	up	to	65°.	The	rotary	setting	allows	for	linkage	
rotation	up	to	95°.	These	settings	only	determine	the	angle	of	sensor	rotation	and	do	not	affect	
control	parameters.	If	a	positioner	is	set	to	linear	linkage	and	a	red	LED	blinks	after	calibration,	the	
most	common	cause	is	that	the	sensor	movement	was	greater	than	65°.	This	can	occur	if	the	roller	
pin	was	placed	in	the	wrong	hole	on	the	follower	arm	or	the	stem	clamp	is	placed	too	high.	The	
take-off	arm	should	always	be	level	with	the	stem	clamp	on	linear	mountings.
The	Logix	3400IQ	positioner	has	an	electrical	measurement	range	of	130°.	That	is,	the	electronics	
will	sense	stem	position	over	a	130°	range	of	travel	of	the	follower	arm.	On	a	rotary	valve,	the	typical	
rotation	is	90°.	When	installing	a	Logix	3400IQ	positioner	on	a	rotary	valve,	the	90°	valve	rotation	
must	be	centered	within	the	130°	electrical	range.	If	mechanical	movement	falls	outside	the	electrical	
measurement	range,	the	positioner	can	have	a	dead	band	at	one	end	of	travel	in	which	valve	move-
ment	cannot	be	sensed.
Question: How do I know if the rotary linkage is centered within the 130° electrical range?
Answer:	The	slot	in	the	take-off	arm	has	enough	clearance	around	the	roller	pin	to	move	the	follower	
arm	slightly.	Move	the	valve	to	the	fully	closed	position.	At	this	position,	move	the	follower	arm	within	
the	slot	clearance.	If	the	valve	does	not	respond	to	the	movement;	linkage	adjustment	is	necessary.	
Repeat	this	test	at	the	fully	open	position.
To	adjust	the	stem	position	linkage,	use	the	A/D	feedback	variable	viewed	using	AD_RAW_FB	param-
eter.	Set	TEST_MODE	bit	‘Enable	diagnostic	Variable	access.’	With	the	valve	in	its	mechanical	fail	
position	(i.e.	no	pressure	applied),	slightly	move	the	follower	arm	while	watching	the	A/D	feedback.	If	
the	number	does	not	change,	the	arm	is	not	centered	in	the	electrical	range.	(The	number	will	bounce	
one	or	two	counts	due	to	noise	at	a	fixed	position	and	should	not	be	considered	a	change,	it	should	
move	greater	than	10	to	20	counts	if	the	linkage	is	centered	correctly).	Rotate	the	take-off	arm,	if	
necessary,	to	bring	the	linkage	in	range.	This	procedure	is	only	necessary	on	a	rotary	mounting.	For	
Linear	mountings,	the	red	LED	will	blink	if	65°	travel	is	exceeded.	Refer	to	the	Calibration	section	for	
further	information	on	stroke	calibration	errors.
Default Configuration 
An	FBAP	containing	default	configuration	parameters	is	resident	in	the	firmware	of	the	device	and	is	
loaded	on	power-up.	By	using	the	NI-FBUS	Configurator	(or	other	fieldbus	configuration)	application,	
the	user	can	create	or	make	changes	to	a	FBAP	for	the	device’s	process	application.
Device Configuration 
Configuring	the	Logix	3400IQ	digital	positioner	results	in:
Function blocks that execute according to a user-defined schedule
Measurements	that	are	processed	according	to	various	user-configured	parameters	found	within	
function	blocks
An output published on the fieldbus network according to a user-defined publishing schedule.
Device Configuration Example
A	sample	printout	of	a	typical	device	configuration	for	the	Logix	3400IQ	digital	positioner	is	given	in	
Appendix	A,	Sample	Configuration	Record.
LAS Capability 
The	Logix	3400IQ	digital	positioner	is	capable	of	operating	as	the	Link	Active	Scheduler	(LAS).	The	
LAS	is	a	fieldbus	device	which	controls	traffic	on	the	network,	such	as	controlling	token-rotation	and	
coordinating	data	publishing.	This	fieldbus	function	is	active	in	only	one	device	at	any	given	time	on	
a	network.	Devices	which	can	be	designated	as	the	LAS	may	be	an	operator	station	or	a	field	device.	
•
•
•
