Triggering, Data acquisition system, Triggering 7.2.3. data acquisition system – Dataman 520 Series User Manual
Page 84
DATAMAN oscilloscope
User’s Guide
Table 7.2.1.1. – Vertical deflection system
7.2.2. Triggering
System type
Dual level
Trigger source for primary
level
selectable Channel A, Channel B or external trigger input
Trigger source for
secondary level
selectable Channel A, Channel B or external trigger input
Threshold setting
Channel A and Channel B on the whole display range.
External fixed on about 1.5V
Slope selection
Leading or trailing edge independently on each source
Minimum trigger pulse
period
5 ns
Minimum trigger pulse
length
2.5 ns
Maximum voltage on
external trigger input
–10V to +13V at 20 kHz or less
Adjustments
Digital filter with ability of setting the valid pulse length
up to 131072*Ts for each level and counter of valid
triggering events selectable from 1 to 32768 for each
level. HOLD-OFF selectable up to 131072*Ts with
selectable AUTO mode, to sample proper amount of data
before trigger. (Ts – actual real time sampling period)
Table 7.2.2.1. – Triggering
7.2.3. Data acquisition system
No of divisions
10
No of pixels per division
50
Mode of operation
Sampling before and after trigger with continual
selection of the trigger position
Record length
8192 samples for each channel
Time base range in 1:1
mode
522
524
526
10 ns/d to 100 ms/d in 1-2-5 sequence
5 ns/d to 100 ms/d in 1-2-5 sequence
2 ns/d to 100 ms/d in 1-2-5 sequence
Time base range using
different ZOOM modes
522
524
526
1 ns/d to 800 ms/d
500 ps/d to 800 ms/d
200 ps/d až 1.6 s/d
Time base accuracy
0.01 % to 100ns/d, 0.5 % for 50ns/d to 5 ns/d
Real time sampling
frequency
522
524
526
500 Hz to 50 MHz
500 Hz to 100 MHz
500 Hz to 200 MHz
Equivalent sampling
frequency
522
524
526
500 Hz to 5 GHz
500 Hz to 10 GHz
500 Hz to 20 GHz
Display range with respect
to trigger event
8190 samples before and 63000 samples after trig. event
in length of 8192 samples
Table 7.2.3.1. – Data acquisition system