DATAMAN oscilloscope

User’s Guide

7.1.3. Data acquisition system

No of divisions

10

No of pixels per division

50

Mode of operation

Sampling before and after trigger with continual
selection of the trigger position

Record length

selectable from 1024 to 1048576 (1M) samples for each
channel

Time base range in 1:1
mode

5 ns/d to 100 ms/d in 1-2-5 sequence

Time base range using
different ZOOM modes

625 ps/d to 204.8 s/d

Time base accuracy

0.01 % to 100ns/d, 0.5 % for 50ns/d to 5 ns/d

Real time sampling
frequency

500 Hz to 100 MHz

Equivalent sampling
frequency

500 Hz to 10 GHz

Display range with respect
to trigger event

1048576 samples before and 1048576 samples after trig.
event in length of 1048576 samples

Table 7.1.3.1. – Data acquisition system

7.1.4. Probe compensation generator

Output connector

BNC, together with External trigger input

Output impedance

1 kOhm to parallel with 10nF and approx. 50 Ohm serial

Output waveform

Pulse with 1:1 duty cycle

Frequency 1465

Hz

Output voltage (no load)

3.3V +- 5%

Table 7.1.4.1. – Compensation generator

7.1.5. Power

Power source

USB interface via USB cable

Max current

USB1.1 – 350mA; USB2.0 – 450mA

Table 7.1.5.1. - Power

7.1.6. Mechanical characteristics

Dimensions without feet
and connectors

165 x 111 x 35 mm

Dimensions with feet and
connectors

182 x 111 x 39 mm

Weight

520 g

Table 7.1.6.1. – Mechanical characteristics

7.1.7. DATAMAN 574 timebase ranges

No REAL

TIME

t/div

1:16

t/div

1:1

t/div

10:1

Max.

ACQ.

TIME

SAMP.

PERIOD

SAMP.

FREQ.

- 82 -